Calibration module not found

Calibration module was not found and the positioning system still needs to be calibrated.

01

Turn on the calibration module by setting the red ON/OFF switch to I. Place the calibration module on the temporary mark. Note the correct orientation
of the calibration module – indicated on the sticker on the calibration module – in relation to the EXCON. The ruler will then be positioned over the black
area on the temporary mark. Make sure there are no obstacles between the calibration module and the radio receivers. Check in Location setup
if the calibration module is found.

The calibration module is foundThe problem is solved.
The calibration module is not foundGo to Step 2
02

Check in Location setup if an unknown tag is detected that only transmits when the calibration module is on.

An unknown tag is visible in Location setup when the calibration module is onGo to Step 3
No unknown tag is visible in Location setup when the calibration module is onGo to Step 4
03

Go to the Maintenance tab and enable the appropriate calibration module.
The serial number of the calibration module can be found on the sticker on the calibration module. Please contact local supplier for instructions.

Check in Location setup if the calibration module is found.

The calibration module is foundThe problem is solved.
The calibration module is not foundGo to Step 4
04

Deploy the spare calibration module. In the Maintenance tab, the deployed calibration module must be enabled.
The serial number of the calibration module can be found on the sticker on the calibration module. Please contact local supplier for instructions.

Check in Location setup if the calibration module is found.

The calibration module is foundThe problem is solved.
The calibration module is not foundPlease contact local supplier.