Calibrating the positioning system fails

01

Make sure there are no obstacles between the calibration module and the radio receivers and check in Location setup whether
the frequency values ​​(Hz) behind the radio receivers do not fluctuate too much.

The frequency values ​​of the radio receivers fluctuate a lotMove the calibration module a few meters and re-measure the field according to the instructions in Training Field Survey. Then go to Step 2
The frequency values ​​of the radio receivers are stableGo to Step 3
02

Connect the laser measurement module with the accompanying USB cable to the USB port on
the EXCON and click on the EXCON on Import laser tool. The measured values ​​are read in.

Green check mark behind the button
All measured values ​​have been read in correctly.
Go to Step 3
Red cross behind the button
The measured values ​​cannot be read correctly.
Go to Importing laser tool does not work.
03

On the EXCON, click Start Calibration again. The system is being calibrated.

Green check mark behind the button
The calibration went well.
The problem is solved.
Red cross behind the button
Calibration did not go well.
Please contact local supplier.